MEDIUM🇬🇧 English

CVE-2017-9657

CVSS 6.5v3.0pub. 2018-04-30upd. 2024-11-21

Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

oryginał EN
CVSS Vector
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
  • Philips Intellivue Mx40

    HW
    Philips
    wszystkie wersje
  • Philips Intellivue Mx40 Firmware

    OS
    Philips
    < b.06.18
🔵
ZWERYFIKUJ U PRODUCENTA
Brak jednoznacznych danych o patchu. Sprawdź referencje od producenta.
CWE
Referencje

Powiązane podatności

CVE-2017-9658MEDIUM6.5ten sam produkt

Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting s...

CVE-2018-8856CRITICAL9.8ten sam vendor

Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The software contains hard-coded cryptograp...

CVE-2018-8850CRITICAL9.8ten sam vendor

Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The software does not validate input proper...

CVE-2017-9656CRITICAL9.1ten sam vendor

The backend database of the Philips DoseWise Portal application versions 1.1.7.333 and 2.1.1.3069 uses hard-co...

CVE-2018-5451CRITICAL9.8ten sam vendor

In Philips Alice 6 System version R8.0.2 or prior, when an actor claims to have a given identity, the software...