Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:HPhilips Intellivue Mx40
HWPhilipswszystkie wersjePhilips Intellivue Mx40 Firmware
OSPhilips< b.06.18
Related vulnerabilities
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06...
Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The software contains hard-coded cryptograp...
Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The software does not validate input proper...
The backend database of the Philips DoseWise Portal application versions 1.1.7.333 and 2.1.1.3069 uses hard-co...
In Philips Alice 6 System version R8.0.2 or prior, when an actor claims to have a given identity, the software...